Model Number V60
Device Problem No Flow (2991)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
Date of event: (b)(6) 2020.Date of report: 09jul2020.
 
Event Description
The patient reported that a flow was not provided.The device was in use with a patient at the time of the issue, however there was no harm reported.The unit was swapped with another unit and therapy continued without patient harm.The manufacturer's international service technician performed a test and check test but the reported phenomenon was not confirmed.The event log was checked as well but no error that indicated airflow arrest was recorded.The manufacturer's international service technician determined that no abnormality was confirmed on the unit however the blower sensor assembly was replaced to be safe.
 
Manufacturer Narrative
G4: 23sep2020.B4: 24sep2020.The reported problem could not be replicated with the returned gas delivery system.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
 
Manufacturer Narrative
G4: 23sep2020; b4: 28sep2020.The reported problem could not be replicated.No fault was found with the returned blower.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V60 VENTILATOR
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
Manufacturer (Section D)
RESPIRONICS CALIFORNIA, INC
2271 cosmos court
carlsbad CA 92011
MDR Report Key10254819
MDR Text Key198737383
Report Number2031642-2020-02301
Device Sequence Number1
Product Code MNT
UDI-Device Identifier00884838033832
UDI-Public(01)00884838033832
Combination Product (y/n)N
PMA/PMN Number
K082660
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Type of Report Initial,Followup,Followup
Report Date 06/16/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/09/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberV60
Device Catalogue Number1076709
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/17/2020
Was the Report Sent to FDA? No
Date Manufacturer Received06/16/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
UNKNOWN PATIENT CIRCUIT, MASK, AND HUMIDIFIER.
-
-