A steris service technician arrived onsite to inspect the 4085 surgical table and found that one of the table's floor locks had a broken foot.This prevented the floor locks from making contact with the floor subsequently causing the table to remain unlocked and the reported event to occur.The root cause of the reported event is attributed to user facility personnel damaging the foot of the floor lock.The 4085 surgical table operator manual states (1-4), "caution - possible equipment damage: when moving the table to or from point of use, roll it carefully at moderate speed and only over smooth floors.Maximum floor clearance is 1/4" (6.4 mm).Avoid door and elevator jambs, and obstructions greater than 1/4" (6.4 mm).If necessary, lift table over obstructions, onto trucks, etc.Lift table evenly and only by the table base." the technician replaced the floor lock, tested the table, confirmed it to be operating according to specifications, and returned it to service.The technician counseled facility personnel on the proper use and operation of the 4085 surgical table, specifically properly moving the table to or from point of use.No additional issues have been reported.
|