• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDIVANCE, INC. ¿ 1725056 ARCTIC SUN 5000 ARCTIC SUN DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDIVANCE, INC. ¿ 1725056 ARCTIC SUN 5000 ARCTIC SUN DEVICE Back to Search Results
Model Number 50000000E
Device Problems Defective Component (2292); Therapeutic or Diagnostic Output Failure (3023)
Patient Problems No Patient Involvement (2645); No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress. Once the investigation is complete a supplemental report will be filed. The device was not returned.
 
Event Description
It was reported that the arctic sun device was not cooling. A functional check showed that the mixing pump had failed and complainant requested a quote for the 2000 hour service.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameARCTIC SUN 5000
Type of DeviceARCTIC SUN DEVICE
Manufacturer (Section D)
MEDIVANCE, INC. ¿ 1725056
321 s taylor ave
louisville CO 80027
Manufacturer (Section G)
MEDIVANCE, INC. ¿ 1725056
321 s taylor ave
louisville CO 80027
Manufacturer Contact
yonic anderson
8195 industrial blvd
covington, GA 30014
7707846100
MDR Report Key10258000
MDR Text Key199062600
Report Number1018233-2020-04444
Device Sequence Number1
Product Code DWJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K161602
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,other,use
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 10/22/2020
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received07/10/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number50000000E
Device Catalogue Number50000000E
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received10/03/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/08/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-