Catalog Number H938739
Device Problems Fluid/Blood Leak (1250); Material Puncture/Hole (1504)
Patient Problem No Patient Involvement (2645)
Event Date 06/11/2020
Event Type  malfunction  
Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that a 1000ml eva (ethyl vinyl acetate) tpn (total parenteral nutrition) bag ¿presented leakage at the front of the bag¿; further described by the reporter as a ¿ hole was evidenced¿.This was identified during setup/preparation prior to patient use.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
Additional information was added to d10, h3, h4 and h6.H4: the lot was manufactured from august 21, 2019 - august 22, 2019.H10: the actual sample was received for evaluation.Unaided visual inspection was performed which observed a tear in the middle front side of the bag and abrasion on the back side of the bag in the same area.A functional testing was performed with water which revealed a leak only from the front side of the bag.Additional magnified inspection verified a tear/hole in front side of the bag where the leak occurred.The reported condition was verified.The cause of the condition could not be determined.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
1000 ML TPN BAG
Type of Device
SET, I.V. FLUID TRANSFER
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
MDR Report Key10259884
MDR Text Key198502473
Report Number1416980-2020-03927
Device Sequence Number1
Product Code LHI
UDI-Device Identifier00085412477237
UDI-Public(01)00085412477237
Combination Product (y/n)Y
PMA/PMN Number
K900585
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Type of Report Initial,Followup
Report Date 08/14/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/10/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/31/2022
Device Catalogue NumberH938739
Device Lot Number60201463
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/04/2020
Date Manufacturer Received08/12/2020
Is This a Reprocessed and Reused Single-Use Device? No