• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WILLIAM COOK EUROPE COOK CELECT PLATINUM NAVALIGN JUGULAR VENA CAVA FILTER SET; DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WILLIAM COOK EUROPE COOK CELECT PLATINUM NAVALIGN JUGULAR VENA CAVA FILTER SET; DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR Back to Search Results
Catalog Number IGTCFS-65-2-JUG-CELECT-PT
Device Problem Difficult or Delayed Activation (2577)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/06/2020
Event Type  malfunction  
Manufacturer Narrative
Manufacturers ref# (b)(4).Blank fields on this form indicate the information is unknown or unavailable.Similar to device under pma/510(k) k171712.Investigation is still in progress.
 
Event Description
Description of event according to initial reporter: physician mentioned that it took 4 attempts to release the ivc filter when ready to deploy.Had released safety button then deployment button in recommended sequence.No adverse effects was reported on the patient due to this occurrence.
 
Event Description
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Manufacturer Narrative
Manufacturers ref# (b)(4).Summary of investigational findings: it took 4 attempts to release the ivc filter when ready to deploy.Had released safety button then deployment button in recommended sequence.No adverse effects to the patient have been reported.According to instruction for use excessive tension during deployment may prevent the filter from releasing when the release mechanism is activated.Based on the provided information it is unknown why the filter was prevented from being released.However, a possible cause is excessive tension during deployment prevented the filter from being released in first attempt.There are adequate controls in place to ensure the device was manufactured to specifications.Cook medical will continue to monitor for similar events.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COOK CELECT PLATINUM NAVALIGN JUGULAR VENA CAVA FILTER SET
Type of Device
DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR
Manufacturer (Section D)
WILLIAM COOK EUROPE
sandet 6
bjaeverskov 4632
DA  4632
MDR Report Key10262952
MDR Text Key200046580
Report Number3002808486-2020-00709
Device Sequence Number1
Product Code DTK
UDI-Device Identifier10827002343105
UDI-Public(01)10827002343105(17)221031(10)E3916408
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,o
Type of Report Initial,Followup
Report Date 09/21/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/13/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/31/2022
Device Catalogue NumberIGTCFS-65-2-JUG-CELECT-PT
Device Lot NumberE3916408
Was Device Available for Evaluation? No
Date Manufacturer Received09/17/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-