• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER CONSTELLATION VISION SYSTEM; UNIT, PHACOFRAGMENTATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER CONSTELLATION VISION SYSTEM; UNIT, PHACOFRAGMENTATION Back to Search Results
Model Number TABLETOP
Device Problem Electrical /Electronic Property Problem (1198)
Patient Problem No Patient Involvement (2645)
Event Date 06/27/2020
Event Type  malfunction  
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is:.
 
Event Description
A biomedical engineer reported low illumination from both ports of the table top illuminator of the system prior to a vitrectomy procedure.The procedure was initiated and completed using the auxiliary illuminator.There was no patient involvement.Additional information was requested and received.
 
Manufacturer Narrative
Additional information has been provided in d.10., h.3., h.6.And h.10.The company service representative examined the system and was able to confirm and replicate the reported event.The company service representative reseated the lamp to address the reported event.The system was then tested and met all product specifications.The system manufacturing device history record (dhr) was reviewed.Based on assessment, the product met specifications at the time of release.The root cause of the reported event can be attributed to the lamp being out of position.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONSTELLATION VISION SYSTEM
Type of Device
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
MDR Report Key10264449
MDR Text Key199732248
Report Number2028159-2020-00519
Device Sequence Number1
Product Code HQC
Combination Product (y/n)N
PMA/PMN Number
K101285
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 10/14/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/13/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/28/2020
Device Model NumberTABLETOP
Device Catalogue Number8065751150
Was Device Available for Evaluation? Yes
Date Manufacturer Received09/25/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-