Catalog Number 04882300001 |
Device Problem
Electrical Shorting (2926)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 06/23/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
The customer¿s products have been requested for return.The product has not been received at this time.If the product is returned in the future, a follow up report will be submitted.
|
|
Event Description
|
The initial reporter had an issue with the accu-chek inform ii meter.He stated the left charging contact appears brown, as if toasted.The base unit appeared to be burnt only on the tip of the left charging contact.The charging contacts on the meter did not appear to be affected.There was no smoke, fire, or injury.
|
|
Manufacturer Narrative
|
Customer's base unit was was received for investigation.Corrosion was observed on one of the charging contacts.The corrosion on the charging contact was caused by contamination due to improper cleaning.Medwatch fields d10 and h3 have been updated.
|
|