Catalog Number SGC0302
Device Problem Leak/Splash (1354)
Patient Problem No Patient Involvement (2645)
Event Date 06/22/2020
Event Type  malfunction  
Manufacturer Narrative
The device was reportedly discarded.Investigation is not yet complete.A follow up report will be submitted with all additional relevant information.
 
Event Description
This is filed to report a leak.It was reported that during preparation of the steerable guide catheter (sgc), the dilator was advanced into the sgc; however, at this time, a loss of fluid column occurred.This occurred three separate times, but on the fourth attempt, no issues occurred.Therefore, the sgc was inserted and two clips were successfully implanted.There was no clinically significant delay in the procedure.No additional information was provided.
 
Manufacturer Narrative
The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other complaints reported from this lot.Based on the information provided a conclusive cause for the reported loss of fluid column could not be determined.There is no indication of a product issue with respect to manufacture, design, or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MITRACLIP SYSTEM STEERABLE GUIDE CATHETER
Type of Device
VALVE REPAIR
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
MDR Report Key10267818
MDR Text Key198876284
Report Number2024168-2020-05732
Device Sequence Number1
Product Code DRA
Combination Product (y/n)N
PMA/PMN Number
K161985
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 08/04/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/13/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/06/2021
Device Catalogue NumberSGC0302
Device Lot Number00403U350
Was Device Available for Evaluation? No
Date Manufacturer Received07/21/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-