MAUDE Adverse Event Report: US SURGICAL PUERTO RICO PROTACK; LAPAROSCOPE, GENERAL & PLASTIC SURGERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

US SURGICAL PUERTO RICO PROTACK; LAPAROSCOPE, GENERAL & PLASTIC SURGERY Back to Search Results
Model Number 174006
Device Problems Loss of or Failure to Bond (1068); Noise, Audible (3273)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/20/2020
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter, during laparoscopic ventral hernia repair procedure, after introducing device into the cavity in order to fix a mesh, the tack didn't fixed correctly and consequently at the peritoneum and an unusual noise was noticed when surgeon pressed the handle.Another device was used however same issue occurred.There was no patient injury.
 
Manufacturer Narrative
Additional information: d10, g4, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted the timing was disrupted.It was reported that the tack did not penetrate the tissue as expected and the device gave unexpected or uncharacteristic audible feedback of normal use.The reported issues were confirmed.The product analysis noted evidence that the device was not used as intended.The issue can occur by an instrument that has been exposed to excessive force while applying helixes to a surface.If a helix is fired over improper surfaces it can provoke the exertion of excessive force to the handle causing the unit to disrupt the timing and to a possible jam.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: do not use the instrument on tissue(s) which cannot be inspected visually for hemostasis.A minimum of 4 mm tissue thickness is required when applying the helical fastener over underlying bone, vessels, or viscera.This device should not be used in tissues that have a direct anatomic relationship to major vascular structures.This would include the deployment of helical fasteners in the diaphragm in the vicinity of the pericardium, aorta or inferior vena cava during diaphragmatic hernia repair.Do not use in ischemic or necrotic tissue.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROTACK
Type of Device
LAPAROSCOPE, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
US SURGICAL PUERTO RICO
201 sabanetas industrial park
ponce PR 00716 4401
MDR Report Key10270561
MDR Text Key198705072
Report Number2647580-2020-02147
Device Sequence Number1
Product Code GCJ
UDI-Device Identifier10884521070400
UDI-Public10884521070400
Combination Product (y/n)N
PMA/PMN Number
K090470
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 08/27/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/14/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number174006
Device Catalogue Number174006
Device Lot NumberP9F1310PY
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/23/2020
Date Manufacturer Received08/11/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-