• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC NEUROMODULATION PRECISION SPECTRA; STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC NEUROMODULATION PRECISION SPECTRA; STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF Back to Search Results
Model Number SC-1132
Device Problems Charging Problem (2892); Intermittent Energy Output (4025)
Patient Problem No Code Available (3191)
Event Date 05/19/2020
Event Type  Injury  
Manufacturer Narrative
Approximated based on the date the manufacturer became aware of the event.
 
Event Description
It was reported that the patient was experiencing intermittent stimulation and the ipg was no longer taking a charge.It was also stated that when the simulation is switched on, it will provide good pain coverage for only an hour.The patient had reprogramming attempts however, the issue still persists.The patient underwent an ipg replacement procedure and was doing well.The explanted ipg was discarded.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRECISION SPECTRA
Type of Device
STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF
Manufacturer (Section D)
BOSTON SCIENTIFIC NEUROMODULATION
25155 rye canyon loop
valencia CA 91355
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
25155 rye canyon loop
valencia CA 91355
Manufacturer Contact
talar tahmasian
25155 rye canyon loop
valencia, CA 91355
6619494863
MDR Report Key10273448
MDR Text Key198856308
Report Number3006630150-2020-02906
Device Sequence Number1
Product Code LGW
UDI-Device Identifier08714729821526
UDI-Public08714729821526
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030017
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 07/14/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/14/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date08/17/2017
Device Model NumberSC-1132
Device Catalogue NumberSC-1132
Device Lot Number18266091
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/25/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/24/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age53 YR
-
-