Device is a combination product.A promus element plus,mr,ous 2.25x32mm stent delivery system was returned for analysis.A visual examination of the stent found stent damage.Microscopic analysis showed evidence of stent damage to most distal stent strut.The undamaged stent outer diameter was measured and the result was within maximum crimped stent profile measurement.The balloon was reviewed, and no issues were noted.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.A visual and tactile examination of the hypotube found multiple kinks.A visual and tactile examination of the shaft polymer extrusion found marker band damage to the distal marker band.An examination of the distal tip found no issues.The device was soaked at 37 degrees because the test guidewire could not track due to residue inside the guidewire lumen.The device was removed from the bath but the recommended 0.014 inch guidewire could not be advanced.No other issues were identified during the product analysis.
|