Catalog Number 8065752438
Device Problem Leak/Splash (1354)
Patient Problem No Patient Involvement (2645)
Event Date 06/26/2020
Event Type  malfunction  
Manufacturer Narrative
On 15-jul-2020, regulatory assessment for reporting complete.With current information, this report of leakage from cassette during calibration meets criteria for reporting based on applicable medical device regulations.The following may apply: exemptions may apply.(b)(6).The manufacturer internal reference number is: (b)(4).
 
Event Description
A customer reported that the cassette leaked before the procedure.The product was replaced and procedure completed with no patient harm.
 
Manufacturer Narrative
A review of the device history record traceable to the reported lot number indicates that the product was processed and released according to the product¿s acceptance criteria.The returned wet sample was visually inspected and there were no obvious defects observed that would have contributed to the reported event.A pressure leak test was then conducted on the cassette utilizing an external pressure source and no cassette leakage was detected.A calibrated constellation console representing a current software version was then used to functionally test the sample.The sample could prime, and pass intraocular pressure (iop) calibration successfully.No anomalies were observed during priming.No system message code was generated during testing.No leakage was detected from the manifolds, connectors, or drain path between the consumable and console.After functional testing no leakage was detected from the pump elastomer or on the pump area of the fluidics module.The presence of fluid leaking from the cassette was not confirmed.After both leakage and console laboratory testing, inspection of the sample indicated no signs of continued leakage from the pump elastomer or cassette.Furthermore, there were no signs of fluid leakage onto the fluidics console due to leakage from the cassette.The cassette passed functional and performance testing.This complaint has been reviewed and the sample was found to be conforming, therefore no further actions will be pursued at this time.Quality assurance has reviewed this complaint and will continue to monitor data for evidence of adverse trending and take further action, as appropriate.Consumables manufacturing management has been made aware of this complaint through the consumer affairs review meeting.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONSTELLATION SURGICAL PROCEDURE PAK
Type of Device
GENERAL SURGERY TRAY (KIT)
Manufacturer (Section D)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
MDR Report Key10276639
MDR Text Key199476716
Report Number1644019-2020-00371
Device Sequence Number1
Product Code LRO
Combination Product (y/n)N
PMA/PMN Number
K880961
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Type of Report Initial,Followup
Report Date 08/27/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/15/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/31/2021
Device Catalogue Number8065752438
Device Lot Number2358227H
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/16/2020
Date Manufacturer Received08/20/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
CONSTELLATION LXT JAPAN
-
-