MAUDE Adverse Event Report: SMITH & NEPHEW, INC. MOTOR DRIVE UNIT HAND CNTRL PWRMX EL; SAW, POWERED, AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. MOTOR DRIVE UNIT HAND CNTRL PWRMX EL; SAW, POWERED, AND ACCESSORIES Back to Search Results
Catalog Number 72200616
Device Problem Failure to Power Up (1476)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/28/2020
Event Type  malfunction  
Manufacturer Narrative
The reported device, used in treatment, was received for evaluation.There was a relationship found between the returned device and the reported incident.A visual inspection was performed on the product and observed no issues.A functional evaluation revealed the blade stalled and the motor was not running.The complaint was confirmed and the root cause has been associated with a mechanical component failure.Factors that could have contributed to the failure include corrosion in the gearbox/motor assembly from cleaning and sterilization methods and the chemicals involved over a period of time or a short in the power cord.No containment or corrective actions are recommended at this time.
 
Event Description
It was reported that during a knee scope, the motor drive unit would not turn on.The procedure was completed without delay using a back-up device.No patient injury or other complications were reported.Results of investigation have concluded that this unit had a blade stall which makes it a reportable event.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Manufacturer Narrative
The reported device, used in treatment, was received for evaluation.There was a relationship found between the returned device and the reported incident.A visual inspection was performed on the product and observed no issues.A functional evaluation revealed the blade stalled and the motor was not running.The complaint was confirmed and the root cause has been associated with a mechanical component failure.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review found related failures.Factors that could have contributed to the failure include corrosion in the gearbox/motor assembly from cleaning and sterilization methods and the chemicals involved over a period of time or a short in the power cord.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MOTOR DRIVE UNIT HAND CNTRL PWRMX EL
Type of Device
SAW, POWERED, AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107
MDR Report Key10280928
MDR Text Key199056402
Report Number1643264-2020-00578
Device Sequence Number1
Product Code HAB
UDI-Device Identifier03596010555472
UDI-Public3596010555472
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Type of Report Initial,Followup
Report Date 07/28/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/15/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number72200616
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/10/2020
Date Manufacturer Received07/24/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-