The customer reported via phone call that they experienced low blood glucose.
Blood glucose reading was 40 mg/dl.
Customer¿s other blood glucose values are 226 mg/dl.
Customer was treated with food for their low.
Customer had symptoms such as shaking, sweating, anxiety, mental confusion, weakness, and fatigue.
The insulin pump will be returned for analysis.
|
Device passed the displacement test, rewind test, prime/seating test, basic occlusion test, force sensor test, occlusion test, sleep current measurement, active current measurement, self test and displacement accuracy test.
A water filled reservoir was used during testing.
Device was programmed with multiple boluses and monitored.
All boluses delivered properly and were listed in the daily history.
Device was programmed with multiple basal profiles and monitored.
All basal profiles delivered their indicated amounts and were verified in the daily and summary history screens.
No delivery anomaly, bolus anomaly or basal anomaly noted during testing.
(b)(4).
|