• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. TAPERLOC STEM; PROSTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. TAPERLOC STEM; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Device Problem Device Contaminated During Manufacture or Shipping (2969)
Patient Problem No Patient Involvement (2645)
Event Date 06/24/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Report source: (b)(6).Product is in process of being returned to zimmer biomet for the investigation.A follow-up mdr will be submitted upon completion.Multiple mdr reports were filed for this event, please see associated reports: 0001825034 - 2020 - 02770, 0001825034 - 2020 - 02771, 0001825034 - 2020 - 02773, 0001825034 - 2020 - 02774, 0001825034 - 2020 - 02775.
 
Event Description
It was reported that during warehouse inspection, debris was found in the sterile packaging.There was no patient involvement.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Complaint sample was returned and evaluated against the reported event.Evaluation of the returned product/photographs provided confirmed the following: debris inside the sterile packaging which is consistent with the appearance of the porous coating and foam debris from the foam packaging inside the sterile barrier.The reported event is confirmed.Dhr was reviewed and no discrepancies were found.The root cause of the reported event is likely to be due to transit damage.The likely condition of the product when it left zimmer biomet was conforming to specification.The device evaluation found no malfunction and the event did not contribute to injury, therefore this would not be considered a reportable event.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TAPERLOC STEM
Type of Device
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key10282987
MDR Text Key199136095
Report Number0001825034-2020-02772
Device Sequence Number1
Product Code KWA
Combination Product (y/n)N
PMA/PMN Number
K110400
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 09/30/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/16/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number51-107140
Device Lot Number3381721
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/24/2020
Was the Report Sent to FDA? No
Date Manufacturer Received09/29/2020
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
-
-