MAUDE Adverse Event Report: MPRI ATTAIN PERFORMA MRI SURESCAN; DRUG ELUTING PERMANENT LEFT VENTRICULAR (LV) PACEMAKER ELECTRODE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MPRI ATTAIN PERFORMA MRI SURESCAN; DRUG ELUTING PERMANENT LEFT VENTRICULAR (LV) PACEMAKER ELECTRODE Back to Search Results
Model Number 429888
Device Problems Adverse Event Without Identified Device or Use Problem (2993); Unstable Capture Threshold (3269)
Patient Problems Muscle Stimulation (1412); Occlusion (1984)
Event Date 06/17/2020
Event Type  Injury  
Manufacturer Narrative
Concomitant medical products: dtma1qq crt-d implanted: (b)(6) 2018.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the patient experienced diaphragmatic stimulation and the left ventricular (lv) lead exhibited unstable thresholds.A lv lead revision was attempted, however, during revision it was noted that the vein was occluded.A venoplasty procedure was performed and the lv lead was removed.A his bundle lead was selected.During implant of the his bundle lead, high thresholds were observed.The his bundle lead was implanted however, shortly after implant the lead dislodged.The his bundle lead remained in use and a revision for the lead was scheduled for a later time.It was also reported that the cardiac resynchronization therapy defibrillator (crt-d) was explanted and replaced due to premature battery depletion that resulted from programmed high output of the lv lead that was needed to maintain capture.No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Product event summary: the full lead was returned and analyzed.No anomalies were found.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTAIN PERFORMA MRI SURESCAN
Type of Device
DRUG ELUTING PERMANENT LEFT VENTRICULAR (LV) PACEMAKER ELECTRODE
Manufacturer (Section D)
MPRI
road 149 km 56.3
villalba PR 00766
MDR Report Key10286051
MDR Text Key199238974
Report Number2649622-2020-13695
Device Sequence Number1
Product Code OJX
UDI-Device Identifier00643169601833
UDI-Public00643169601833
Combination Product (y/n)N
PMA/PMN Number
P080006
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 07/29/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/16/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date09/10/2019
Device Model Number429888
Device Catalogue Number429888
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/07/2020
Date Manufacturer Received07/17/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
6935M55 LEAD, 5076-45 LEAD
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age72 YR
Patient Weight91
-
-