Investigation evaluation: the product said to be involved was returned in an open pouch from the lot number provided in the report.The label matches the product returned.Our laboratory evaluation of the product said to be involved confirmed the report and determined the drive wire had detached from the handle.The device was returned with the basket fully retracted into the sheath.The handle was manipulated and no response from the basket was noted.The handle was disassembled and it was noted that the drive wire had detached from the handle and nesting was observed in the purple hub.For further evaluation of the drive wire cable and basket, the catheter was cut to push the drive wire cable out of the sheath.The basket was fully formed and intact, and evidence of solder was present on the handle cannula at the joint.Liquid was observed inside of the catheter and an unknown yellow substance was observed on the distal tip of the basket.No other anomalies were detected with the devices.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: a definitive cause for this observation could not be determined because the actual use conditions could not be duplicated in the laboratory setting.Due to a variety of clinical conditions such as patient anatomy, endoscope position or progression of disease state, we could not reproduce the actual conditions of product usage during our laboratory analysis.This limits our ability to conclusively determine a cause.The instructions for use indicates: "advance device through channel, in short increments, until basket sheath exits endoscope." the instructions for use also states: "confirm desired position of basket sheath relative to target.Advance basket out of sheath.Caution: pulling on sheath while advancing or retracting basket may damage device, rendering it inoperable." basket deployment difficulties and buckling/breaking of the drive wire can occur if the device experiences excessive pressure.Resistance in basket movement and bends in the catheter can occur if the elevator of the endoscope is used to deflect the device at a sharp angle.Prior to distribution, all memory ii double lumen extraction baskets are subjected to a visual inspection and functional testing to ensure device integrity.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
|
During an endoscopic procedure, the physician used the memory ii double lumen extraction basket (mb-35-2x4-8) in the bile duct.After the first stone removal, the user pulled the basket into the catheter and then tried to open the basket but it did not come out from the sheath.Another device was used instead.There was no reportable information at this time.The device was received for evaluation on (b)(6) 2020 where it was determined that the drive wire had detached from the handle subject of report.A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
|