Model Number CI-1500-04 |
Device Problems
Device Slipped (1584); Intermittent Communication Failure (4038)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 06/24/2020 |
Event Type
malfunction
|
Event Description
|
The recipient is reportedly experiencing loss of lock.Device testing could not completed due to loss of lock.External equipment was exchanged, however, the issue did not resolve.Revision surgery will be scheduled.
|
|
Manufacturer Narrative
|
The recipient's device was reportedly explanted.The recipient was reimplanted with another advanced bionics cochlear device.
|
|
Manufacturer Narrative
|
Advanced bionics considers the investigation into this reportable event as closed.The device passed the external visual and photographic imaging inspections.System lock was obtained intermittently.The electrode condition prevented an electrical test from being performed.The scanning electron microscopy analysis revealed cracked epoxy at the feedthru pin connection on some pins.The device passed the mechanical tests performed.This device malfunctioned due to a crack in the conductive epoxy at the feedthru pin-to-substrate connection.A corrective action was implemented.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|