ARTHROCARE CORP. DISPOSABLE FIRSTPASS SUTURE PASSER, SELF Back to Search Results
Catalog Number 22-4038
Device Problem Mechanical Problem (1384)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/29/2020
Event Type  malfunction  
Event Description
It was reported that during a shoulder procedure, the handle of the firstpass was stiff and could not be depressed in order to pass the needle.The procedure was successfully completed without significant delay using a back-up device.No other complications were reported.
 
Manufacturer Narrative
H10.H3, h6: the device reported, used in treatment, was not returned for evaluation.A relationship between the product and reported incident cannot be established.A review of manufacturing records for the reported lot number found no non-conformances or anomalies during manufacturing process related to the reported event.A complaint history review found no related failures; this failure mode will be trended to assess for any necessary corrective actions.A visual inspection and functional evaluation cannot be performed and customer´s complaint cannot be confirmed.Potential factors unrelated to the design or manufacture of the device that may lead to the failure reported include, but are not limited to: (1) excessive force (2) tissue thickness.(3) damage or debris in the device tip between passes.No containment or corrective actions are recommended at this time.There were no indications during manufacturing record review that would suggest that the device did not meet product specifications upon release into distribution.H11.Corrected information in d4 (udi number), g5 (pma/510(k)number) and h1 (type of reportable event).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DISPOSABLE FIRSTPASS SUTURE PASSER, SELF
Type of Device
PASSER
Manufacturer (Section D)
ARTHROCARE CORP.
7000 w. william cannon
austin TX 78735
MDR Report Key10293083
MDR Text Key199448311
Report Number3006524618-2020-00514
Device Sequence Number1
Product Code HWQ
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 09/18/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/18/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date11/28/2022
Device Catalogue Number22-4038
Device Lot Number2044997
Was Device Available for Evaluation? No
Date Manufacturer Received09/16/2020
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-