DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID TYPE B PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
|
Back to Search Results |
|
Model Number N/A |
Device Problem
Connection Problem (2900)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 06/29/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
A getinge service territory manager (stm) instructed the customer via telephone to replace the helium tank with a known full helium tank.The customer then verified that helium was displayed on iabp screen and verified that the issue was resolved.
|
|
Event Description
|
It was reported that during a routine check performed by the customer, the cardiosave intra-aortic balloon pump (iabp) was not recognizing the new helium tank.There was no patient involved and no adverse event was reported.
|
|
Manufacturer Narrative
|
Updated fields: b4, g4, g7, h2, h6(evaluation method codes), h10.
|
|
Event Description
|
It was reported that during a routine check performed by the customer, the cardiosave intra-aortic balloon pump (iabp) was not recognizing the new helium tank.There was no patient involved and no adverse event was reported.
|
|