Model Number N/A
Device Problem Connection Problem (2900)
Patient Problem No Patient Involvement (2645)
Event Date 06/29/2020
Event Type  malfunction  
Manufacturer Narrative
A getinge service territory manager (stm) instructed the customer via telephone to replace the helium tank with a known full helium tank. The customer then verified that helium was displayed on iabp screen and verified that the issue was resolved.
 
Event Description
It was reported that during a routine check performed by the customer, the cardiosave intra-aortic balloon pump (iabp) was not recognizing the new helium tank. There was no patient involved and no adverse event was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameCARDIOSAVE HYBRID TYPE B PLUG
Type of DeviceSYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ 07430
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ 07430
Manufacturer Contact
1300 macarthur blvd
mahwah, NJ 07430
MDR Report Key10302020
MDR Text Key199696385
Report Number2249723-2020-01118
Device Sequence Number1
Product Code DSP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 07/29/2020
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received07/21/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberN/A
Device Catalogue Number0998-00-0800-53
Device Lot NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received07/29/2020
Was Device Evaluated by Manufacturer? No
Date Device Manufactured06/01/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-