• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW MEDICAL LTD. RENASYS TOUCH DEVICE & POWER SUP; NEGATIVE PRESSURE WOUND THERAPY POWERED SUCTION PUMP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW MEDICAL LTD. RENASYS TOUCH DEVICE & POWER SUP; NEGATIVE PRESSURE WOUND THERAPY POWERED SUCTION PUMP Back to Search Results
Catalog Number 66801280
Device Problem Electrical Power Problem (2925)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/26/2020
Event Type  malfunction  
Event Description
It was reported that during service evaluation j13 on the main board was found broken therefore battery cannot be recharged also the device cannot start up correctly.No case involved.
 
Manufacturer Narrative
H10.H3, h6: the device, intended for use in treatment, was returned for evaluation.The visual inspection found defects to the outer case.The functional evaluation found that the device could not operate on battery or mains power and could not be charged, due to this it could not switch on, establishing a relationship with the event reported.It was established that the root cause of this was a failed j13 component on the main circuit board.A review of manufacturing records for the reported lot/batch, found no non-conformances or anomalies during production.The device/product met all specifications upon release into distribution.Complaint history for the reported event has been reviewed, revealing further instances in the past three years.No further actions are deemed necessary at this stage.However, we will continue to monitor for any adverse trends relating to this product range.Smith + nephew acknowledge customer concern and are continually investigating ways to develop and improve our products.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RENASYS TOUCH DEVICE & POWER SUP
Type of Device
NEGATIVE PRESSURE WOUND THERAPY POWERED SUCTION PUMP
Manufacturer (Section D)
SMITH & NEPHEW MEDICAL LTD.
101 hessle road
hull east riding of yorkshire HU3 2 BN
UK  HU3 2BN
MDR Report Key10305325
MDR Text Key199823007
Report Number8043484-2020-02143
Device Sequence Number1
Product Code OMP
Combination Product (y/n)N
PMA/PMN Number
K153209
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Type of Report Initial,Followup
Report Date 10/25/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/21/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number66801280
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/26/2020
Date Manufacturer Received10/23/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-