RESPIRONICS CALIFORNIA, INC V60 VENTILATOR VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
|
Back to Search Results |
|
Model Number V60 |
Device Problem
Incorrect, Inadequate or Imprecise Resultor Readings (1535)
|
Patient Problem
No Patient Involvement (2645)
|
Event Type
Malfunction
|
Manufacturer Narrative
|
Date of event: (b)(6) 2020.
Date of report: 21jul2020.
|
|
Event Description
|
The customer reported that the unit alarmed with a check vent machine pressure sensor autozero error failure while on a test lung.
The customer contacted product support and was advised to turn and off the unit to see if the problem will returned.
Per the service manual, product support recommend replacing the solenoid valve and provided the part number.
The customer reported that the unit was not in use on a patient.
The biomedical engineer said that he could not duplicate the reported problem.
He said that he turned on and of f the unit all day and the reported error did not returned.
The unit is now back in service.
|
|