Model Number 5392
Device Problem Calibration Problem (2890)
Patient Problem No Patient Involvement (2645)
Event Date 06/26/2020
Event Type  malfunction  
Manufacturer Narrative
Product analysis: analysis was able to confirm the customer comment of the rate control not working.Analysis noted that the encoder assembly was out of electrical specification, atrial output connector was broken, hanger was bent and cracked, keypad window was scratched, one control knob was contaminated, lower case was broken, main label was damaged, main seal was damaged and missing section by battery drawer, all six plugs were damaged, need battery shortening bar, both display wires were pinched (not compromised), both wires in lower case were pinched (not compromised) and wires were incorrectly routed over the battery compartment.Analysis noted that the epg had evidence of a non-manufacturer service person disassembling and reassembling the epg.The main seal was damaged, missing section by battery drawer, six plugs were damaged with tool marks, both display wires were pinched and both battery wires in lower case were pinched.All found defective parts were replaced and all other identified issues were resolved.The device passed all final functional tests.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the external pulse generator ( epg) knob was not working.It was noted that during a rate change the dial would turn but the pacing rate did not change in the 78-80 range.The device was returned for service.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DUAL CHAMBER TEMPORARY PACEMAKER
Type of Device
PULSE-GENERATOR, PACEMAKER, EXTERNAL
Manufacturer (Section D)
PLEXUS MANUFACTURING SDN. BHD
bayan lepas free industrial zo
bayan lepas 11900
MY  11900
Manufacturer (Section G)
PLEXUS MANUFACTURING SDN. BHD
bayan lepas free industrial zo
bayan lepas 11900
MY   11900
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key10313478
MDR Text Key200034671
Report Number3004593495-2020-00771
Device Sequence Number1
Product Code DTE
UDI-Device Identifier00643169230859
UDI-Public00643169230859
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K150246
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 07/23/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/23/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number5392
Device Catalogue Number5392
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/02/2020
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/26/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/20/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-