Model Number XS PST |
Device Problem
Image Display Error/Artifact (1304)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 07/03/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
The customer¿s products have been requested for return.The product has not been received at this time.If the product is returned in the future, a follow up report will be submitted.Occupation was lay user/patient.
|
|
Event Description
|
The initial reporter had an issue with the display of the coaguchek xs meter that may result in a misinterpretation of results.During a display check, the customer's wife stated she only saw dashes and the greater than symbol on the display.The display is missing segments.No actual results have been misinterpreted.
|
|
Manufacturer Narrative
|
The customer's meter was received for investigation.The meter¿s circuit board was tested for damage or contamination and it was found the printed circuit board (pcb) was contaminated.The root cause is determined to be contamination of the contacts due to improper handling or maintenance by the customer.Medwatch fields d10 and h3 were updated.
|
|