Model Number NM-3138-55 |
Device Problem
Defective Device (2588)
|
Patient Problem
No Code Available (3191)
|
Event Date 06/30/2020 |
Event Type
Injury
|
Manufacturer Narrative
|
Additional suspect medical device components involved in the event: model: nm-3138-55, serial/ lot: (b)(4), description: 55cm 8 contact extension.
|
|
Event Description
|
It was reported that following the implantation of the extensions, one of four was tested and did not function as intended.A revision was performed in which the extension was removed and replaced.The patient has recovered since the procedure.
|
|
Event Description
|
It was reported that following the implantation of the extensions, one of four was tested and did not function as intended.A revision was performed in which the extension was removed and replaced.The patient has recovered since the procedure.
|
|
Manufacturer Narrative
|
The lead nm-3138-55 serial number: (b)(6) was returned and analyzed.The allegations of the lead not functioning as intended have been confirmed.X-ray inspection of the lead extension revealed four cables were fractured after the weld in the distal connector stack.This type of damage typically occurs when excessive tensile force is exerted onto the lead body and connector section of the lead.Bending the distal end could also cause cable fractures in the connector section of the lead.The broken cables are still contained inside the connector.
|
|
Search Alerts/Recalls
|