Udi: (b)(4) stenosis of an implanted valve may be a manifestation of structural valve deterioration (svd).This term refers to changes intrinsic to the valve, and can include failure modes such as wear, calcification, leaflet tear, stent creep, leaflet disruption, or leaflet retraction.There are cases of svd that result in a combination of regurgitation and stenosis.It may be mild and not require any intervention or it may be moderate to severe.In these cases, it causes the heart to work harder to eject blood from the ventricle.Depending on severity it could be an indication for valve replacement or medical intervention.It is possible patient factors such as metabolic issues contributed to the valve stenosis.A very common failure mode is tissue calcification.The mechanisms for bioprosthetic heart valve tissue calcification are not fully understood.Many factors can contribute to the onset and propagation of calcification including patient related (e.G.Patient age, disease state, immune status, and other co-morbidities), pharmacological, and intrinsic properties of the valve itself.It is widely understood that patients with chronic renal disease and prior history of calcific stenosis of the native valve may be predisposed to bioprosthetic calcification.A dhr review was performed and did not reveal any issues that may have contributed to the complaint event.Per the instructions for use (ifu), paravalvular leak (pvl) and valve regurgitation are potential adverse events associated with bioprosthetic heart valves and tavr procedure.Paravalvular leak refers to blood flowing through a channel between the structure of the implanted valve and the cardiac tissue, as a result of a lack of appropriate sealing of the valve to the target site.Some pvl is not uncommon post deployment.Many cases are mild to moderate, and either resolve over time or do not cause symptoms.Others may be more clinically significant and require intervention.The mechanism behind worsening or late pvl is not well understood but may be related to cardiac remodeling.Valve regurgitation which develops progressively over time can be due to a number of issues including patient related factors or structural valve deterioration, including calcification, non-calcific degeneration, leaflet thickening or fibrosis, or a combination of these.Regurgitation may also develop progressively if host fibrotic tissue, or pannus, grows onto the bioprosthetic valve.Pannus, a cause of nonstructural dysfunction, may interfere with functionality of the device by restricting the leaflet motion leading to abnormal coaptation.In this case, the exact cause of the stenosis, worsening pvl, and working valve regurgitation could not be confirmed.However, it is possible patient factors may have contributed to the events.A review of edwards lifesciences risk management documentation was performed for this case.The reported event is an anticipated risk of the transcatheter heart valve procedure, additional assessment of the failure mode is not required at this time.The ifu and training manuals have been reviewed and no inadequacies have been identified with regards to warnings, contraindications, and the directions/conditions for the successful use of the device.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis, and any excursions above the control limits are assessed and documented as part of this monthly review.No corrective or preventative actions are required.
|
As reported, approximately 3 years post successful tf tavr a valve in valve was performed, implanting a 23mm s3u valve within the pre-existing 23mm s3 valve.Based on additional information received, approximately 2.5 years post successful tavr the patient experienced increased velocity across the valve and increased gradient.Moderate prosthetic regurgitation was also noted.Prior to the valve in valve procedure, 2 separate jets of pvl were observed along with mildly reduced leaflet opening.Around the area of the left coronary cusp there was an eccentric jet of pvl that was moderate and a central jet that was also moderate.
|