COOK INC GUNTHER TULIP FEMORAL VENA CAVA FILTER SET DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR Back to Search Results
Model Number G33016
Device Problems Difficult to Remove (1528); Device Tipped Over (2589)
Patient Problems Vessel Or Plaque, Device Embedded In (1204); Dyspnea (1816); Pain (1994)
Event Type  Injury  
Manufacturer Narrative

The previous medwatch report was submitted by william cook europe under manufacturer report reference#3002808486-2020-00512. Additional information provided determined that this device was manufactured by cook inc. With the submission of this initial medwatch report, cook inc. Informs that all future submissions regarding this complaint will be handled under the manufacturer report reference # blank fields on this form indicate the information is unknown, unavailable, or unchanged. (b)(6). Investigation the following allegations have been investigated: embedment, unable to remove, tilt, pain, shortness of breath (sob), limited mobility. Investigation is reopened due to additional information provided. The reported allegations have been further investigated based on the information provided to date. Filter interacts with ivc wall, e. G. Penetration/perforation/embedment. This may be either symptomatic or asymptomatic. Potential causes may include improper deployment; and (or) excessive force or manipulations near an in-situ filter (e. G. , a surgical or endovascular procedure in the vicinity of a filter). Potential adverse events that may occur include, but are not limited to, the following: trauma to adjacent structures, vascular trauma, vena cava perforation, vena cava penetration. Physician practice guidelines and published guidance from regulatory agencies recommend that patients with indwelling filters undergo routine follow-up. The risks/benefits of filter retrieval should be considered for each patient during follow-up. Once protection from pe is no longer necessary, filter retrieval should be considered. Filter retrieval should be attempted when feasible and clinically indicated. Filter retrieval is a patient-specific, clinically complex decision; the decision to remove a filter should be based on each patient¿s individual risk/benefit profile (e. G. , a patient¿s continued need for protection from pe compared to their experience with and (or) ongoing risk of experiencing filter-related complications). For all retrievable ivc filters, retrieval becomes more challenging with time, and this is commonly due to encapsulation of the filter legs or hook (in a tilted filter) by tissue ingrowth. The filter is designed to be retrieved with the günther tulip vena cava filter retrieval set. It may also be retrieved with the cloversnare® vascular retriever. Cook has not performed testing to evaluate the safety or effectiveness of filter retrieval using other retrieval systems or techniques. The published clinical literature includes descriptions of alternative techniques for filter retrieval; use of these techniques varies according to physician experience, patient anatomy, and filter position. The safety or effectiveness of these alternative retrieval techniques has not been established. Specific for ¿embedded¿ a filter that is embedded in the wall of the ivc may be difficult to retrieve. For all retrievable ivc filters, retrieval becomes more challenging with time, and this is commonly due to encapsulation of the filter legs or hook (in a tilted filter) by tissue ingrowth. Filter tilt has been reported. Potential causes may include filter placement in ivcs with diameters larger than those specified in these instructions for use; improper deployment; manipulations near an implanted filter (e. G. , a surgical or endovascular procedure in the vicinity of a filter); and (or) a failed retrieval attempt. Excessive filter tilt may contribute to difficult or failed retrieval; vena cava wall penetration/perforation; and (or) result in loss of filter efficiency. Potential adverse events that may occur include, but are not limited to, the following: unacceptable filter tilt. Unknown if the reported pain, shortness of breath and limited mobility are directly related to the filter and unable to identify a corresponding failure mode at this point in time. 20 devices in lot. No relevant notes on work order. The product is manufactured and inspected according to specifications. No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e. G. Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava. This report includes information known at this time. A follow-up medwatch report will be submitted if additional relevant information becomes available.

 
Event Description

Patient allegedly received an implant on (b)(6) 2012 via the right common femoral artery due to deep vein thrombosis. The patient alleges tilt, device is unable to be retrieved, and embedment. The patient further alleges pain, shortness of breath, limited mobility. (b)(6) 2017, per a report from computed tomography; ¿impression: gunther tulip inferior vena cava filter in place, unchanged in position without evidence of filter migration. Ivc retrieval hook is again apposed to the inferior vena cava wall consistent with an embedded filter. ¿.

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameGUNTHER TULIP FEMORAL VENA CAVA FILTER SET
Type of DeviceDTK FILTER, INTRAVASCULAR, CARDIOVASCULAR
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer (Section G)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer Contact
larry pool
750 daniels way
bloomington, IN 47404
8123392235
MDR Report Key10320149
MDR Text Key201906464
Report Number1820334-2020-01357
Device Sequence Number1
Product Code DTK
Combination Product (Y/N)N
Reporter Country CodeUS
PMA/PMN NumberK043509
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type CONSUMER
Reporter Occupation
Type of Report Initial
Report Date 07/24/2020
1 Device Was Involved in the Event
1 Patient Was Involved in the Event
Date FDA Received07/24/2020
Is This An Adverse Event Report? Yes
Is This A Product Problem Report? Yes
Device Operator HEALTH PROFESSIONAL
Device EXPIRATION Date06/01/2015
Device MODEL NumberG33016
Device Catalogue NumberIGTCFS-65-FEM
Device LOT Number3466038
Was Device Available For Evaluation? No
Is The Reporter A Health Professional?
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received07/14/2020
Was Device Evaluated By Manufacturer? Device Not Returned To Manufacturer
Date Device Manufactured06/01/2012
Is The Device Single Use? Yes
Is this a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient TREATMENT DATA
Date Received: 07/24/2020 Patient Sequence Number: 1
-
-