OLYMPUS MEDICAL SYSTEMS CORP. ENDOSCOPE REPROCESSOR Back to Search Results
Model Number OER-PRO
Device Problems Defective Component (2292); Component Missing (2306)
Patient Problem No Patient Involvement (2645)
Event Date 06/16/2020
Event Type  malfunction  
Manufacturer Narrative
The device was not returned for evaluation.The field service engineer onsite repaired the device by replacing the grey connectors.Once completed, the device was tested and passed all required testing and specifications.
 
Event Description
It was reported that the grey air/water/instrument channel connector on the oer-pro was found broken and the valve was missing.There were no other details provided regarding the event.There was no patient involvement on this report.
 
Manufacturer Narrative
The device history records for this device were reviewed and all records indicated that the product was manufactured according to all applicable procedures and met final product release criteria.No abnormalities were found.The cause of the event cannot be conclusively determined.Probable cause could be due to some impact that caused damage (chipped) to the grey connector, or the connector unit was loose and the unit came apart.Possible causes include stress to the device unit during cleaning.Design of the device or manufacturing, cannot be confirmed as factors to cause of the event.Ifu(instructions for use) states: check the following for each connector.The connector should be fixed firmly.The o-rings should be free of abnormalities such as cracks, tears, or dents.Do not use the equipment in any connector seems to be damaged or defective.Using the equipment when an irregularity has been detected may interfere with reprocessing.Furthermore, fluid leakage may damage peripheral devices or facilities near the equipment.Olympus will continue to monitor complaints for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOSCOPE REPROCESSOR
Type of Device
ENDOSCOPE REPROCESSOR
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
MDR Report Key10323380
MDR Text Key223253133
Report Number8010047-2020-04786
Device Sequence Number1
Product Code FEB
UDI-Device Identifier04953170258589
UDI-Public04953170258589
Combination Product (y/n)N
PMA/PMN Number
K103264
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 10/27/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/24/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberOER-PRO
Was the Report Sent to FDA? No
Date Manufacturer Received09/30/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-