The device history records for this device were reviewed and all records indicated that the product was manufactured according to all applicable procedures and met final product release criteria.No abnormalities were found.The cause of the event cannot be conclusively determined.Probable cause could be due to some impact that caused damage (chipped) to the grey connector, or the connector unit was loose and the unit came apart.Possible causes include stress to the device unit during cleaning.Design of the device or manufacturing, cannot be confirmed as factors to cause of the event.Ifu(instructions for use) states: check the following for each connector.The connector should be fixed firmly.The o-rings should be free of abnormalities such as cracks, tears, or dents.Do not use the equipment in any connector seems to be damaged or defective.Using the equipment when an irregularity has been detected may interfere with reprocessing.Furthermore, fluid leakage may damage peripheral devices or facilities near the equipment.Olympus will continue to monitor complaints for this device.
|