Model Number 97725
Device Problems High impedance (1291); Shipping Damage or Problem (1570)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/22/2020
Event Type  malfunction  
Manufacturer Narrative
Concomitant medical products: product id 977d260, serial# (b)(4).Product type screening device product id 977d260 lot# serial# (b)(4).Product type screening device.Other relevant device(s) are: product id: 977d260, serial/lot #: (b)(4), ubd: 21-aug-2023, udi#: (b)(4).If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received by a manufacturer representative (rep) regarding a patient with an implantable neurostimulator (ins).It was reported that the impedance measurements were 40k on multiple contacts prior to use.The lead connectivity showed bad connection.The lead was repositioned in the wens a few times and impedances were wrong.The wens was changed and connectivity was wrong.The lead was then replaced and a new lead/wens resolved the issue.The wens and lead with the high impedance were expected to be returned for analysis.No symptoms or further complications were reported.
 
Manufacturer Narrative
H3: analysis of product id: 977d260, serial# (b)(6), found no anomaly.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTELLIS
Type of Device
STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF
Manufacturer (Section D)
MEDTRONIC NEUROMODULATION
800 53rd ave ne
minneapolis MN 55421 1200
MDR Report Key10332449
MDR Text Key200629563
Report Number3007566237-2020-00346
Device Sequence Number1
Product Code LGW
UDI-Device Identifier00763000017590
UDI-Public00763000017590
Combination Product (y/n)N
PMA/PMN Number
P840001
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 10/21/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/28/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/14/2021
Device Model Number97725
Device Catalogue Number97725
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/05/2020
Date Manufacturer Received10/20/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age76 YR
-
-