Model Number N/A |
Device Problem
Power Problem (3010)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 07/06/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
A getinge field service engineer (fse) was dispatched to evaluate this unit.The fse confirmed the reported issue.The fse replaced the power supply and motor controller board.The iabp was returned for clinical use.
|
|
Event Description
|
It was reported that the cs300 intra-aortic balloon pump (iabp) would not switch on.A second power supply was tried but also was not working.There was no patient involvement, and no adverse event reported.
|
|
Manufacturer Narrative
|
Updated fields: b4, g4, g7, h2, h6 (evaluation method codes), h10.
|
|
Event Description
|
It was reported that the cs300 intra-aortic balloon pump (iabp) would not switch on.A second power supply was tried but also was not working.There was no patient involvement, and no adverse event reported.
|
|