Model Number XS PST |
Device Problem
Image Display Error/Artifact (1304)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 07/01/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
The reporter's product was requested for investigation and replacement product was sent to the reporter.Occupation - the occupation is lay user/patient.
|
|
Event Description
|
The initial reporter stated there was an issue with the display of coaguchek xs meter serial number (b)(4).The reporter had just replaced the batteries to the meter with 4 new duracell batteries.The reporter stated she turned the meter on and saw all items on the display and it was all flashing.The meter beeped multiple times and did not proceed to test mode.The reporter removed the batteries and the battery compartment was clean and dry, with no corrosion.The reporter cleaned the battery contacts of the meter with a pencil eraser.The reporter then performed a display check of the meter and segments were missing from the results field of the display.
|
|
Manufacturer Narrative
|
The customer's meter was received for investigation.The display shows no error during investigation.The meter¿s circuit board was tested for damage or contamination and it was found the printed circuit board (pcb) was contaminated.The root cause is determined to be contamination of the contacts due to improper handling or maintenance by the customer.Medwatch fields d9 and h3 were updated.
|
|