• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS COAGUCHEK XS SYSTEM; PROTHROMBIN TIME MONITOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS COAGUCHEK XS SYSTEM; PROTHROMBIN TIME MONITOR Back to Search Results
Model Number XS PST
Device Problem Image Display Error/Artifact (1304)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/01/2020
Event Type  malfunction  
Manufacturer Narrative
The reporter's product was requested for investigation and replacement product was sent to the reporter.Occupation - the occupation is lay user/patient.
 
Event Description
The initial reporter stated there was an issue with the display of coaguchek xs meter serial number (b)(4).The reporter had just replaced the batteries to the meter with 4 new duracell batteries.The reporter stated she turned the meter on and saw all items on the display and it was all flashing.The meter beeped multiple times and did not proceed to test mode.The reporter removed the batteries and the battery compartment was clean and dry, with no corrosion.The reporter cleaned the battery contacts of the meter with a pencil eraser.The reporter then performed a display check of the meter and segments were missing from the results field of the display.
 
Manufacturer Narrative
The customer's meter was received for investigation.The display shows no error during investigation.The meter¿s circuit board was tested for damage or contamination and it was found the printed circuit board (pcb) was contaminated.The root cause is determined to be contamination of the contacts due to improper handling or maintenance by the customer.Medwatch fields d9 and h3 were updated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COAGUCHEK XS SYSTEM
Type of Device
PROTHROMBIN TIME MONITOR
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
MDR Report Key10335432
MDR Text Key200801294
Report Number1823260-2020-01805
Device Sequence Number1
Product Code GJS
Combination Product (y/n)N
PMA/PMN Number
K062925
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Type of Report Initial,Followup
Report Date 10/01/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/28/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberXS PST
Device Catalogue Number03666794001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/17/2020
Date Manufacturer Received07/01/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-