Model Number DL900F
Device Problems Malposition of Device (2616); Patient Device Interaction Problem (4001)
Patient Problem Pain (1994)
Event Type  Injury  
Manufacturer Narrative

Manufacturing review: a lot history review was performed. This is the only complaint to date for this lot number. Therefore, a device history record review is not required. Investigation summary: the device was not returned for evaluation. The medical records included images. Images and medical records were provided and reviewed. Based on the image review, the ivc filter was tilted anteriorly 25 degrees. Grade 1 ivc leg extension beyond the wall of the ivc with some extension into a right sided lumbar vein. No grade 2 interactions. Based on the image review, the investigation is confirmed for the perforation of the ivc and filter tilt. Based on the medical record review, bard denali filter was implanted at the vertebral interspace clearly below the right renal vein for a patient with left lower extremity deep vein thrombosis. However, no device deficiencies were identified in the provided medical record. Based on the medical record review, the investigation is inconclusive for the alleged perforation of the ivc and filter tilt as no objective evidence has been provided to confirm any alleged deficiency with the filter. Therefore, the investigation is confirmed for the perforation of the inferior vena cava (ivc) and filter tilt. Based on the available information, the definitive root cause is unknown. Labeling review: a review of product labeling documents (e. G. Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, and unit label) showed that the product labeling is adequate. (expiry date: 09/2017).

 
Event Description

It was reported through the litigation process that a vena cava filter was placed in a patient after being diagnosed with deep vein thrombosis/pulmonary embolism. At some time post filter deployment, it was alleged that the filter perforated into lumbar veins. The device has not been removed and there were no reported attempts made to retrieve the filter. The patient reportedly experienced lower back pain; however, the current status of the patient is unknown.

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameDENALI FEMORAL SYSTEM
Type of DeviceVENA CAVA FILTER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
Manufacturer (Section G)
C.R. BARD, INC. (GFO)
289 bay road
queensbury NY 12804
Manufacturer Contact
judith ludwig
1415 w. 3rd street
tempe, AZ 85281
4803032689
MDR Report Key10342609
MDR Text Key200961175
Report Number2020394-2020-05211
Device Sequence Number1
Product Code DTK
Combination Product (Y/N)N
Reporter Country CodeUS
PMA/PMN NumberK130366
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type CONSUMER,HEALTH PROFESSIONAL
Reporter Occupation
Type of Report Initial
Report Date 07/30/2020
1 Device Was Involved in the Event
1 Patient Was Involved in the Event
Date FDA Received07/30/2020
Is This An Adverse Event Report? Yes
Is This A Product Problem Report? Yes
Device Operator HEALTH PROFESSIONAL
Device MODEL NumberDL900F
Device Catalogue NumberDL900F
Device LOT NumberGFYH3668
Was Device Available For Evaluation? No
Is The Reporter A Health Professional?
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received07/08/2020
Was Device Evaluated By Manufacturer? Device Not Returned To Manufacturer
Is The Device Single Use? Yes
Is this a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient TREATMENT DATA
Date Received: 07/30/2020 Patient Sequence Number: 1
-
-