MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. 630G INSULIN PUMP MMT-1715KL 630G ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. 630G INSULIN PUMP MMT-1715KL 630G ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND Back to Search Results
Model Number MMT-1715KL
Device Problem Material Integrity Problem (2978)
Patient Problems Fatigue (1849); Hyperglycemia (1905)
Event Date 07/24/2020
Event Type  Injury  
Manufacturer Narrative

(b)(4). Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned. The device will be returned for analysis and further information will follow once the analysis has been completed. No conclusion can be drawn at this time.

 
Event Description

The customer reported via phone call that the customer experienced high blood glucose levels. The customer¿s blood glucose levels were 430 mg/dl and 496 mg/dl. The customer had been using insulin pump system within 48 hours of reported high blood glucose event. The customer reported that they did not feel well, felt tired and hungry. The customer treated them with correction bolus. The customer also reported battery cap issue but the high blood glucose event continued even after battery change. The insulin pump will not be returned for analysis.

 
Manufacturer Narrative

Device passed displacement test, rewind test, prime/seating test, basic occlusion test, force sensor test, occlusion test, sleep current measurement, active current measurement, selftest and displacement accuracy test. Device passed the delivery accuracy test. No battery cap received with unit. (b)(4).

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name630G INSULIN PUMP MMT-1715KL 630G
Type of DeviceARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer Contact
gerwin de graaff
ceiba norte ind. park #50 road
juncos 00777--386
8185464805
MDR Report Key10342943
MDR Text Key200966897
Report Number2032227-2020-155072
Device Sequence Number1
Product Code OZO
Combination Product (Y/N)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type CONSUMER
Reporter Occupation
Type of Report Initial,Followup
Report Date 08/31/2020
1 Device Was Involved in the Event
1 Patient Was Involved in the Event
Date FDA Received07/30/2020
Is This An Adverse Event Report? Yes
Is This A Product Problem Report? No
Device Operator LAY USER/PATIENT
Device MODEL NumberMMT-1715KL
Device Catalogue NumberMMT-1715KL
Device LOT NumberHG3JXUC
Was Device Available For Evaluation? Device Returned To Manufacturer
Date Returned to Manufacturer08/11/2020
Is The Reporter A Health Professional? No
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received08/31/2020
Was Device Evaluated By Manufacturer? Device Not Returned To Manufacturer
Date Device Manufactured06/27/2019
Is The Device Single Use? No
Is this a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unkown

Patient TREATMENT DATA
Date Received: 07/30/2020 Patient Sequence Number: 1
Treatment
FRN-UNK-RSVR, UNOMED INF SET
-
-