• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL - NEUROMODULATION OCTRODE LEAD KIT, 60CM LENGTH; PERCUTANEOUS LEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL - NEUROMODULATION OCTRODE LEAD KIT, 60CM LENGTH; PERCUTANEOUS LEAD Back to Search Results
Model Number 3186
Device Problems Fracture (1260); High impedance (1291)
Patient Problem Inadequate Pain Relief (2388)
Event Date 07/01/2020
Event Type  Injury  
Manufacturer Narrative
(b)(6).Date of event is estimated.Further information was requested but not received.Unique device identifier (udi #): the udi is unknown because the lot number was not provided.The results/method and conclusion codes along with investigation results will be provided in the final report.
 
Event Description
It was reported that the patient lost therapy due to high impedances.Reportedly, the lead is fractured.As such, surgical intervention may take place at a later date to address the issue.
 
Event Description
Related manufacturer reference number: 1627487-2020-31763 additional information received indicated surgical intervention took place on (b)(6) 2020 wherein the leads were explanted and replaced with new leads addressing the issue.Reportedly, therapy was confirmed post operatively.New information revealed that patient had 2 leads.
 
Manufacturer Narrative
The results of the investigation are inconclusive, since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OCTRODE LEAD KIT, 60CM LENGTH
Type of Device
PERCUTANEOUS LEAD
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
MDR Report Key10347046
MDR Text Key201358379
Report Number1627487-2020-23423
Device Sequence Number1
Product Code LGW
Combination Product (y/n)N
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup,Followup
Report Date 02/11/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/30/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number3186
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received02/10/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
SCS LEAD, MODEL 3186
Patient Outcome(s) Other;
-
-