Catalog Number 1500350-18
Device Problems Break (1069); Leak/Splash (1354)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/08/2020
Event Type  malfunction  
Manufacturer Narrative
The device was received.Investigation is not yet complete.A follow report will be submitted with all additional relevant information.
 
Event Description
It was reported that the procedure was performed to treat an unspecified lesion.The 3.50x18mm rx xience sierra leaked from the hub during inflation.It was then noticed that the hub was cracked.The device was replaced with a new xience to successfully complete the procedure.There were no adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.
 
Manufacturer Narrative
A visual and functional inspection was performed on the returned sds.The reported hub cracks and leak were confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents and/or complaints reported from this lot.The investigation was unable to determine a conclusive cause for the reported hub cracks resulting in leaks.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIENCE SIERRA EVEROLIMUS ELUTING CORONARY STENT SYSTEM
Type of Device
DRUG ELUTING CORONARY STENT DELIVERY SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
MDR Report Key10349712
MDR Text Key202071559
Report Number2024168-2020-06353
Device Sequence Number1
Product Code NIQ
Combination Product (y/n)Y
PMA/PMN Number
P110019
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 09/02/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/31/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/15/2022
Device Catalogue Number1500350-18
Device Lot Number9070941
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/28/2020
Date Manufacturer Received08/10/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
INDEFLATOR
-
-