It was reported that the patient lost therapy due to ipg being inoperable in (b)(6) 2020.The ipg was not communicating with external devices.As such, surgical intervention took place on (b)(6) 2020 wherein the ipg was explanted and replaced with a new ipg addressing the issue.Reportedly, therapy was restored post operatively.
|
The reported event of inoperable ipg was confirmed.As received, the ipg battery was depleted.The battery was recovered and was able to communicate with a lab patient programmer.However, the ipg would not charge with the lab charging system.Troubleshooting revealed that the inability to charge was isolated to components q15 and q16.Q15 and q16 are fets in the primary regulator circuit that switch on and off to modulate the coil current.Since q15 and q16 were degraded, the absorption of rf energy from the charger was limited and vchg could not maintain sufficient voltage to enable charging.The degraded components q15 and q16 were isolated to charging only and had no effect on stimulation output or communication with a patient programmer.
|