NIHON KOHDEN CORPORATION GF-210RA; MULTI-GAS UNIT Back to Search Results
Model Number GF-210RA
Device Problem Output Problem (3005)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/08/2020
Event Type  malfunction  
Manufacturer Narrative
The customer reported that the multigas unit is displaying an "external device" error message.No harm or injury was reported.The customer will be sending the failed unit in for an exchange.Nihon kohden continues to investigate the reported event.Nihon kohden will submit a supplemental report in accordance with 21 cfr section 803.56 when additional information becomes available.The following fields contain no information (ni), as attempts to obtain information were made, but not provided: concomitant medical products.
 
Event Description
The customer reported that the multigas unit is displaying an "external device" error message.
 
Manufacturer Narrative
Details of complaint: on (b)(6) 2020 customer stated gas unit did not gives "external device failure" on the host monitor.The failure is related to gas sensor.Investigation summary the root cause of reported issue was determined to be firmware related.An update is required to correct this issue.Capa (b)(4) has been initiated.
 
Event Description
The customer reported that the multigas unit is displaying an "external device" error message.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GF-210RA
Type of Device
MULTI-GAS UNIT
Manufacturer (Section D)
NIHON KOHDEN CORPORATION
1-31-4 nishiochia, shinjuku-ku
attn: shama mooman
tokyo, 161-8 560
JA  161-8560
MDR Report Key10350555
MDR Text Key201572470
Report Number8030229-2020-00422
Device Sequence Number1
Product Code CCK
UDI-Device Identifier04931921106891
UDI-Public04931921106891
Combination Product (y/n)N
PMA/PMN Number
K110594
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Type of Report Initial,Followup
Report Date 08/11/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/31/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberGF-210RA
Device Catalogue NumberGF-210RA
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/20/2020
Was the Report Sent to FDA? Yes
Date Report Sent to FDA08/11/2020
Distributor Facility Aware Date08/07/2020
Device Age21 MO
Event Location Hospital
Date Report to Manufacturer08/11/2020
Date Manufacturer Received08/07/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-