MAUDE Adverse Event Report: MICRO THERAPEUTICS, INC. DBA EV3 PIPELINE FLEX LASER THERAPY PRODUCT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MICRO THERAPEUTICS, INC. DBA EV3 PIPELINE FLEX LASER THERAPY PRODUCT Back to Search Results
Model Number PED-500-35
Device Problems Activation Failure (3270); Physical Resistance/Sticking (4012)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/27/2020
Event Type  malfunction  
Manufacturer Narrative
Since the device was not returned, we are unable to perform further root cause analysis and the exact cause of the reported event is unknown. All devices are 100% tested and all products are 100% inspected for damages and irregularities during manufacture. If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received report that the pipeline braid distal tip kinked and could not be opened. The device was removed from the treating vessel. The device finally was able to be opened when force was applied. More than 50% of the device was deployed when it failed to open. More than 2 resheathing attempts were made. This event occurred during the treatment of a left, unruptured, saccular aneurysm. The max diameter was 12. 1 mm and the neck was 2. 3 mm. The distal landing zone was 3. 24 mm and the proximal was 4. 98 mm. The vessel anatomy was moderate in tortuosity. No patient injury occurred. The devices were prepared and used per the instructions for use (ifu). Partial resistance, within normal range, was felt when the ped was advanced. A continuous flush was used.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NamePIPELINE FLEX
Type of DeviceLASER THERAPY PRODUCT
Manufacturer (Section D)
MICRO THERAPEUTICS, INC. DBA EV3
9775 toledo way
irvine CA 92618
Manufacturer (Section G)
MICRO THERAPEUTICS, INC. DBA EV3
9775 toledo way
irvine CA 92618
Manufacturer Contact
tricha miles
9775 toledo way
irvine, CA 92618
7635140379
MDR Report Key10350678
MDR Text Key203120308
Report Number2029214-2020-00758
Device Sequence Number1
Product Code OUT
UDI-Device Identifier00847536016903
UDI-Public00847536016903
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
P100018.S011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation
Type of Report Initial,Followup,Followup
Report Date 10/27/2020
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received07/31/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date07/25/2022
Device Model NumberPED-500-35
Device Catalogue NumberPED-500-35
Device Lot NumberA871380
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/11/2020
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received10/27/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/26/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-