Catalog Number 0684-00-0497
Device Problem Leak/Splash (1354)
Patient Problems Pain (1994); No Consequences Or Impact To Patient (2199)
Event Date 07/09/2020
Event Type  Injury  
Manufacturer Narrative
The product has been returned to the manufacturer, but is pending investigation. Once the investigation is completed a supplemental report with our findings will be submitted. Complaint # (b)(4).
 
Event Description
It was reported that during intra-aortic balloon (iab) therapy, there was a "leak in circuit" alarm generated and blood visible in the iab tubing. The insertion was reported to be axillary, which is not the method described in the device instructions for use. The iab was replaced and therapy was provided. There was no reported injury to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameMEGA 8FR. 50CC IAB
Type of DeviceSYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - FAIRFIELD
15 law drive
fairfield NJ 07004
Manufacturer (Section G)
DATASCOPE CORP. - FAIRFIELD
15 law drive
fairfield NJ 07004
Manufacturer Contact
dorota wolpiuk
15 law drive
fairfield, NJ 07004
MDR Report Key10354977
MDR Text Key201398709
Report Number2248146-2020-00386
Device Sequence Number1
Product Code DSP
Combination Product (y/n)N
PMA/PMN Number
K091449
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation
Type of Report Initial,Followup,Followup
Report Date 10/12/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/03/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator
Device Catalogue Number0684-00-0497
Device Lot Number3000111228
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer07/16/2020
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Event Location No Information
Date Manufacturer Received10/09/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/20/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage

Patient Treatment Data
Date Received: 08/03/2020 Patient Sequence Number: 1
-
-