RESPIRONICS CALIFORNIA, INC V60 VENTILATOR; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
|
Back to Search Results |
|
Model Number V60 |
Device Problem
Failure to Zero (1683)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Date of event: (b)(6) 2020.Date of report: 03aug2020.
|
|
Event Description
|
The customer reported the machine pressure sensor autozero failed.The device was in clinical use at the time of the issue, however there was no patient harm reported.The manufacturer's field service engineer (fse) was unable to duplicate the problem during evaluation, but the error did appear in the significant event log.The fse replaced the solenoid valve and the issue was resolved.
|
|
Manufacturer Narrative
|
G4: 13jul2020.B4: 04aug2020.D11: concomitant medical products updated.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
|
|