• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC, INC. PACING SYSTEM ANALYZER; PROGRAMMER, PACEMAKER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC, INC. PACING SYSTEM ANALYZER; PROGRAMMER, PACEMAKER Back to Search Results
Model Number 229047
Device Problem Connection Problem (2900)
Patient Problem No Patient Involvement (2645)
Event Date 07/06/2020
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the analyzer connection with the programmer was not working properly , the connector was faulty.The analyzer was returned for analysis.There was no patient involvement reported.
 
Manufacturer Narrative
Product analysis: analysis was able to confirm the customer comment that the analyzer connection with the programmer was not working properly.It was noted that the analyzer failed the incoming functional test at several tests.It was further noted that the battery was depleted.Additionally the device was noted as being mechanically intact.Due to the parts required to repair the analyzer not being available, the device has been scrapped.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PACING SYSTEM ANALYZER
Type of Device
PROGRAMMER, PACEMAKER
Manufacturer (Section D)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
MDR Report Key10360930
MDR Text Key201514506
Report Number2182208-2020-01417
Device Sequence Number1
Product Code KRG
UDI-Device Identifier00721902256011
UDI-Public00721902256011
Combination Product (y/n)N
PMA/PMN Number
P890003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Type of Report Initial,Followup
Report Date 08/21/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/04/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number229047
Device Catalogue Number229047
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/08/2020
Date Manufacturer Received08/13/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-