• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIVANOVA DEUTSCHLAND HEATER-COOLER SYSTEM 3T CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA DEUTSCHLAND HEATER-COOLER SYSTEM 3T CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 16-02-80
Device Problem Microbial Contamination of Device (2303)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/01/2020
Event Type  malfunction  
Manufacturer Narrative
There was no known patient involvement. The heater-cooler 16-02-80 is not distributed in the usa, but it is similar to heater-cooler 16-02-85, which is distributed in the usa (510(k) number: k191402). Livanova (b)(4) implemented a field safety notice for disinfection and cleaning of heater-cooler devices. The z number is z-2076/2081-2015. Livanova (b)(4) manufactures the heater-cooler system 3t. The incident occurred in (b)(6). Through follow-up communication livanova learned that the device was cleaned regularly per the instruction for use and placed outside the operation theatre during use. The device will be sent to the manufacturer site for deep disinfection in order to solve the reported issue. If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
Livanova (b)(4) received a report that a heater-cooler system 3t device was found to be contaminated with mycobacterium chimaera. The laboratory report confirming the reported contamination has been provided to livanova. There is no known patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameHEATER-COOLER SYSTEM 3T
Type of DeviceCONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80309
GM 80309
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80309
GM 80309
Manufacturer Contact
enrico greco
14401 w. 65th way
arvada, CO 80004
MDR Report Key10362490
MDR Text Key201889493
Report Number9611109-2020-00451
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Reporter Country CodeIT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other Health Care Professional
Remedial Action Recall
Type of Report Initial
Report Date 07/06/2020
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received08/04/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number16-02-80
Was Device Available for Evaluation? No Answer Provided
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Event Location No Information
Date Manufacturer Received07/06/2020
Was Device Evaluated by Manufacturer? No Answer Provided
Date Device Manufactured01/31/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberZ-2076/2081-2015

-
-