Model Number 71674001 |
Device Problem
Defective Device (2588)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 07/05/2020 |
Event Type
Injury
|
Event Description
|
It was reported that the surgeon has been having trouble with getting the distal screw in a short nail.Always wants to go posterior to the nail.In order to complete the procedure, a change in surgical technique was required.Any other issue that could affect the patient's condition was not reported by this event.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
|
Manufacturer Narrative
|
The device, used in treatment, was returned for evaluation.A visual inspection of the returned drill guide handle did not find any obvious issues.The device shows normal signs of wear/usage.The device was manufactured in 2013.A functional evaluation was conducted with a mating device and the stated failure was confirmed.The drill guide handle will not hold a mating device as intended.A medical investigation will be performed.Proceed based on information provided/available for the investigation; if no relevant clinical information is provided, recommend closure.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of risk management files found that the reported failure was documented appropriately.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Expected wear/tear is likely the probable cause of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.
|
|
Search Alerts/Recalls
|
|
|