MAUDE Adverse Event Report: CORDIS CORPORATION UNKNOWN OPTEASE VENA CAVA FILTER FILTER, INTRAVASCULAR, CARDIOVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CORDIS CORPORATION UNKNOWN OPTEASE VENA CAVA FILTER FILTER, INTRAVASCULAR, CARDIOVASCULAR Back to Search Results
Model Number N/A
Device Problems Filter (816); Failure to Align (2522)
Patient Problem Perforation (2001)
Event Date 07/21/2020
Event Type  Injury  
Manufacturer Narrative

Please note that device reported is an optease/trapease vena cava filter and for which the catalog and lot numbers are not currently available. Patient age and medical history were also not provided. If obtained, a follow up report will be submitted within 30 days upon receipt. Please note that the contact is not a medical professional but a more accurate choice is not available. It was reported that a patient underwent placement of an optease vena cava filter. The information provided indicated that the filter subsequently malfunctioned and caused tilt and perforation of the inferior vena cava (ivc) wall. The indication for the filter implant, procedural details and medical history of the patient have not been provided and there is currently no additional information available for review. The product was not returned for analysis and the sterile lot number has not been provided; therefore, no device analysis nor device history record review could be performed. The optease vena cava filter is indicated for use in the prevention of recurrent pulmonary embolism (pe) via percutaneous placement in the vena cava for patients in which anticoagulants are contraindicated, anticoagulant therapy for thromboembolic disease has failed, emergency treatment following massive pe where anticipated benefits of conventional therapy are reduced or for chronic, recurrent pe where anticoagulant therapy has failed, or is contraindicated. The purpose of a vena cava filter is to catch thrombus from the lower extremities as it travels along normal blood flow patterns up towards the heart. Ivc filter tilt has been associated with the anatomy of the vessel, specifically asymmetry and tortuosity. Vessel perforation is a known adverse event associated with implanting vena cava filters and is listed as such in the instructions for use (ifu). The ifu also notes vessel damage such as intimal tears and perforation as procedural and long-term complications related to ivc filters. The timing and mechanism of the tilt and perforation has not been reported at this time and a clinical conclusion could not be determined as to the cause of the event. It is unknown if the tilt contributed to the reported perforation. Without images available for review the reported events could not be confirmed or further clarified. Given the limited information available for review at this time, there is nothing to suggest that the reported events are related to the design and manufacturing process of the device; therefore, no corrective action will be taken. Should additional information become available, the file will be updated accordingly.

 
Event Description

As reported in the legal brief, a patient was treated with an optease vena cava filter which subsequently malfunctioned and caused injury, damage, to the patient, including, but not limited to specific evidence that the filter was tilted and perforated the ivc wall. As a direct and proximate result of these malfunctions, the patient suffered life-threatening injuries and damages, and required extensive medical care and treatment. As a further proximate result, the patient has suffered and will continue to suffer significant medical expenses, and pain and suffering, and other damages.

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameUNKNOWN OPTEASE VENA CAVA FILTER
Type of DeviceFILTER, INTRAVASCULAR, CARDIOVASCULAR
Manufacturer (Section D)
CORDIS CORPORATION
14201 nw 60th ave
miami lakes FL 33014
Manufacturer (Section G)
CORDIS CORPORATION
14201 nw 60th ave
miami lakes FL 33014
Manufacturer Contact
karla castro
14201 nw 60th ave
miami lakes, FL 33014
7863138372
MDR Report Key10368976
MDR Text Key201896758
Report Number1016427-2020-04257
Device Sequence Number1
Product Code DTK
Combination Product (Y/N)N
Reporter Country CodeUS
PMA/PMN NumberK034050
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type OTHER
Reporter Occupation OTHER HEALTH CARE PROFESSIONAL
Type of Report Initial
Report Date 08/05/2020
1 Device Was Involved in the Event
1 Patient Was Involved in the Event
Date FDA Received08/05/2020
Is This An Adverse Event Report? Yes
Is This A Product Problem Report? No
Device Operator HEALTH PROFESSIONAL
Device MODEL NumberN/A
Device Catalogue Number466FXXXX
Was Device Available For Evaluation? Yes
Is The Reporter A Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received07/27/2020
Was Device Evaluated By Manufacturer? No
Is The Device Single Use? Yes
Is this a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient TREATMENT DATA
Date Received: 08/05/2020 Patient Sequence Number: 1
-
-