U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD ACCESS SYSTEMS ACCUCATH INTRAVASCULAR CATHETER 20 GX2.25 CATHETER,INTRAVASCULAR,THERAPEUTIC,SHORT-TERM LESS THAN 30 DAYS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD ACCESS SYSTEMS ACCUCATH INTRAVASCULAR CATHETER 20 GX2.25 CATHETER,INTRAVASCULAR,THERAPEUTIC,SHORT-TERM LESS THAN 30 DAYS Back to Search Results
Model Number N/A
Device Problem Device Dislodged or Dislocated (2923)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
The device has not been returned to the manufacturer for evaluation. A lot history review (lhr) of reep1702 showed no other similar product complaint(s) from this lot number.
 
Event Description
It was reported by the rn "i had attempted to place an accucath #20. I advanced the guidewire without issue, and advanced the catheter without issue as well. Once i pressed the needle release button, it would not retract. I had attempted to adjust my angle but still would not retract. I had to pull back slightly and there was resistance, but the needle finally retracted. I tried flushing the catheter and it had become dislodged after retracting the needle. This has happened about 3 times with the #20 accucath. When i pulled the catheter from the patient, it looked slightly kinked. ".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameACCUCATH INTRAVASCULAR CATHETER 20 GX2.25
Type of DeviceCATHETER,INTRAVASCULAR,THERAPEUTIC,SHORT-TERM LESS THAN 30 DAYS
Manufacturer (Section D)
BARD ACCESS SYSTEMS
605 n. 5600 w.
salt lake city UT 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas
MX
Manufacturer Contact
kelsey erickson
605 n. 5600 w.
salt lake city, UT 84116
8015225937
MDR Report Key10370000
MDR Text Key201765186
Report Number3006260740-2020-02770
Device Sequence Number1
Product Code FOZ
UDI-Device Identifier00801741110948
UDI-Public(01)00801741110948
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K153298
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation
Type of Report Initial
Report Date 08/05/2020
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received08/05/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date06/30/2021
Device Model NumberN/A
Device Catalogue NumberAC0202250
Device Lot NumberREEP1702
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location Hospital
Date Manufacturer Received07/13/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/01/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-