A steris service technician arrived onsite to inspect the 3085 sp surgical table.The technician tested the function and articulations of the table and found the table to be operating properly; no issues were identified, and the table was returned to service.During the time of the reported event, the table was tilted and in trendelenburg position with a safety strap on the patient's legs.While onsite, the technician spoke with user facility personnel who indicated that the safety strap should have been properly secured during the time of the reported event.The reported event is attributed to user error as user facility personnel should have ensured the patient was properly secured.The 3085 sp surgical table operator manual states (1-2), "patient must be secured to the table in accordance with recommended positioning practices." a steris account manager performed in-service training on the proper use and operation of the table, specifically ensuring the patient is properly secured.No additional issues have been reported.
|