• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EDWARDS LIFESCIENCES EDWARDS SAPIEN 3 TRANSCATHETER HEATH VALVE; AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EDWARDS LIFESCIENCES EDWARDS SAPIEN 3 TRANSCATHETER HEATH VALVE; AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY Back to Search Results
Model Number 9600TFX23J
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Complete Heart Block (2627); No Code Available (3191)
Event Date 06/03/2020
Event Type  Injury  
Manufacturer Narrative
Udi: (b)(4).Per the instructions for use (ifu), conduction system defects (heart block) which may require a permanent pacemaker are potential adverse events associated with balloon aortic valvuloplasty, deployment of the prosthetic valve, and the overall tavr procedure.According to the valve academic research consortium (varc) guidelines, the close anatomical relationship between the aortic valve complex and the branching atrioventricular bundle may provide an explanation for these complications of the tavr procedure.According to literature review, and as documented in a technical summary written by edwards lifesciences ¿clinical technical summary for complaints-conduction disturbances/ heart block¿, atrioventricular conduction disturbances after tavr are associated with many patient related and procedural related factors, including pre-operative co-morbid status, the degree and bulkiness of aortic valve and annular calcification, inter-ventricular septal thickness, pre-existing electrocardiogram abnormalities, the depth of prosthesis implantation, and the profile of the implanted prosthesis.Unlike conventional avr, where there may be localized trauma due to decalcification of the annulus and/or suture placement in the proximity of the av node or the bundles, tavr may cause conduction abnormalities through mechanical impingement of the conduction system by the prosthesis.The mechanisms of the development of heart block after tavr are well documented and described in the literature.It is also documented that pre-existing heart block is common in patients undergoing tavr or surgical avr and another 4-6 % will develop postoperative heart block, potentially requiring a permanent pacemaker.In this case, there was no allegation or indication a device malfunction contributed to this adverse event.The device was not returned for evaluation, as it remained implanted.The root cause of this event remains indeterminable but was likely due to patient related factors and/or the mechanisms described above.The ifu and training manuals have been reviewed and no inadequacies have been identified with regards to warnings, contraindications, and the directions/conditions for the successful use of the device.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis, and any excursions above the control limits are assessed and documented as part of this monthly review.No corrective or preventative actions are required at this time.
 
Event Description
As reported through the (b)(6) tavi case registry, complete atrioventricular (av) block was observed on postoperative day (pod) 5 of the transfemoral tavr procedure for a 23mm transcatheter valve.On pod 13, a permanent pacemaker was implanted and the patient was discharged from the hospital.The outcome was determined as in remission.Per medical opinion, the event was not associated with the edwards device and the tavr procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EDWARDS SAPIEN 3 TRANSCATHETER HEATH VALVE
Type of Device
AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY
Manufacturer (Section D)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
Manufacturer (Section G)
EDWARDS LIFESCIENCES LLC
1 edwards way
irvine CA 92614
Manufacturer Contact
rene dorne
1 edwards way
linc 4-1
irvine, CA 92614
9492506385
MDR Report Key10376780
MDR Text Key202710557
Report Number2015691-2020-12959
Device Sequence Number1
Product Code NPT
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
P140031
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 07/12/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/06/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date02/01/2022
Device Model Number9600TFX23J
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/23/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/20/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-