Model Number 100014514
Device Problems Failure of Device to Self-Test (2937); Output Problem (3005)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/05/2020
Event Type  Injury  
Manufacturer Narrative
The results, method and conclusion codes along with investigation results will be provided in the final report.
 
Event Description
During an ablation procedure, after ablation had been started the error message "amplifier error, log out and power cycle both the ensite velocity dws and amplifier before restarting. Contact sjm technical support" was displayed and the amplifier status light turned orange. The systems were power cycled and upon restarting ablation the same issue occurred. Troubleshooting was not able to resolve the issue and the procedure was cancelled. There were no adverse consequences to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameENSITE VELOCITY SYSTEM VELOCITY AMPLIFIER
Type of DeviceCOMPUTER, DIAGNOSTIC, PROGRAMMABLE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.
one st. jude medical drive
st. paul MN 55117
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.
one st. jude medical drive
st. paul MN 55117
Manufacturer Contact
stephanie o' sullivan
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key10378249
MDR Text Key202004331
Report Number2184149-2020-00120
Device Sequence Number1
Product Code DQK
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K141050
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation
Type of Report Initial,Followup
Report Date 09/14/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/07/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator
Device Model Number100014514
Device Catalogue Number100014514
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/11/2020
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Event Location No Information
Date Manufacturer Received09/10/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

Patient Treatment Data
Date Received: 08/07/2020 Patient Sequence Number: 1
-
-