Model Number 16-02-80
Device Problem Unexpected Shutdown (4019)
Patient Problem No Patient Involvement (2645)
Event Date 07/09/2020
Event Type  Malfunction  
Manufacturer Narrative

There was no patient involvement. Livanova (b)(4) manufactures the heater-cooler system 3t. The incident occurred in (b)(6). Livanova field service representative was dispatched to the facility and confirmed the issue. The reported failure was traced back to a damaged cooling compressor. The root cause is a hole of the evaporator with consequent leak of coolant and water entering the refrigeration cycle. This led to compressor damages. The unit will be removed from service and scrapped.

 
Event Description

Livanova received a report that the heater-cooler system 3t tripped the mains power at the start up. There was no patient involvement.

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameHEATER-COOLER SYSTEM 3T
Type of DeviceCONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80309
GM 80309
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80309
GM 80309
Manufacturer Contact
enrico greco
14401 w. 65th way
arvada, CO 80004
MDR Report Key10378673
MDR Text Key202105358
Report Number9611109-2020-00453
Device Sequence Number1
Product Code DWC
Combination Product (Y/N)N
Reporter Country CodeFR
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type HEALTH PROFESSIONAL
Reporter Occupation OTHER HEALTH CARE PROFESSIONAL
Remedial Action Other
Type of Report Initial
Report Date 08/06/2020
1 Device Was Involved in the Event
0 PatientS WERE Involved in the Event:
Date FDA Received08/07/2020
Is This An Adverse Event Report? No
Is This A Product Problem Report? Yes
Device Operator HEALTH PROFESSIONAL
Device MODEL Number16-02-80
Was Device Available For Evaluation? No Answer Provided
Is The Reporter A Health Professional? Yes
Was the Report Sent to FDA? No
Event Location No Information
Date Manufacturer Received07/09/2020
Was Device Evaluated By Manufacturer? Yes
Date Device Manufactured10/28/2015
Is The Device Single Use? No
Is this a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-