The device was returned because the distal end of the scope was broken and exposed there is more information on the device evaluation.This supplemental report is being submitted to provide this information.The device history record review confirmed that device has no abnormalities, special adoption, or variations in manufacturing.There is no repair history for the device.The likely cause of the tip damage is because some external force was applied to the device.The instructions for use includes the following statements in chapter 6 compatible reprocessing methods and chemical agents: attach the sterilization cap to the endoscope before ethylene oxide gas sterilization.If the sterilization cap is not attached to the endoscope during ethylene oxide gas sterilization, the air inside the endoscope will expand and could rupture the bending section cover and/or damage the angulation mechanism (see figure 6.2).Exceeding the recommended parameters may cause equipment damage.Do not exceed the parameters outlined in tables 6.3 and 6.4 on page 63.Do not seal the endoscope with other equipment that may contact the endoscope in the package.Otherwise, the endoscope may be damaged.
|